Abstract

Introduction: Around half of world blindness and fifth of visual impairment are caused by cataract. Treatment of cataract and  refractive errors correction improve 73.6% of blindness. Preservation of the Anterior chamber and corneal endothelium are very important aims during cataract surgery. Methodology: This prospective, observational study was done at King Fahd Hospital of the University (KFUH), Alkhobar, Saudi Arabia. After getting the Institutional Review Board (IRB) approval, the study was initiated at KFUH between April 2016 and August 2016. After getting the patients' informed consents, a thorough assessment was done to each patient prior to surgery. All eyes enrolled in the study have undergone phacoemulsification surgery procedure. For each patient, one month after the phacoemulsification surgery , specular microscopy was performed. The phacodynamic parameters and surgery time were compared with the degree of endothelial cells loss, hexagonality and central corneal thickness. Results: The phacodynamic parameters used during phacoemulsification, namely, power (AVG %), ultrasound time, and elliptical motion showed a strong correlation with the change of endothelial cell density (all P< .001). The length of surgery also showed a positive correlation with the amount of endothelial cell loss ( P < .001). All of these parameters did not significantly affect corneal endothelial hexagonality except elliptical motion. Change in central corneal thickness was very minimal. Conclusion: Endothelial loss was significantly associated with phacoemulsification power, ultrasound time, elliptical motion and length of surgery.

Keywords: Phacoemulsification,Cataract,Corneal endothelium,Specular microscopy,Hexagonality,Endothelial cells

Received: 15 March 2017 / Revised: 3 May 2017 / Accepted:29 May 2017/ Published: 13 June 2017

Contribution/ Originality

This study is one of very few studies which have investigated the effect of phacoemulsification cataract surgery on density and shape of corneal endothelial cells.


1. INTRODUCTION

The most common surgery done for American population above 65 years is cataract surgery [1] . Cataract is the most common cause of impaired vision in the world [2]. The prevalence of blindness in Saudi Arabia is 0.7% (52.6% due to cataract) and the prevalence of vision impairment is 10.9% (20.6% due to cataract). Treatment of cataract and refractive errors correction help 73.6% of blind subjects and 88.5% of visually handicapped population [3]. Recently, Cataract surgery has taken giant steps of improvement including new techniques, new instruments and the use of viscoelastics [4]. Phacoemulsification has become the most common cataract procedure since its introduction in 1967 [5] Preservation of the eye’s anatomy is a very important aim during cataract surgery so that patients can get the best benefit from cataract extraction [5, 6]. More concern is related to the protection of corneal endothelium [7-9] because this single layer of cells which plays an important role in preservation of corneal transparency [10] doesn't regenerate once damaged [11, 12]. Corneal Endothelial cell damage is mediated through prostaglandins which lead to significant inflammation and corneal swelling and increased thickness [13]. Factors that lead to corneal endothelial injury include the jackhammering effect of phaco cutting, implosion of air bubbles in the anterior chamber, the ultrasonic power fluctuation, bouncing of nuclear fragments, fluid turbulence and production of free radicals [14-17]. Irrigation pressure and temperature rise play their roles as well in harming corneal endothelial cells [18]. Surgeon's experience, hardness of cataract, type of corneal incision, age and preoperative visual acuity influenced endothelial cell loss as well [19]. Central endothelial cell count decreases by an average rate of 0.6% each year throughout life [20]. Specular is very useful tools for evaluation of corneal endothelium, with confocal microscopy showing more detailes about corneal endothelial histocytology [21, 22]. Corneal Endothelial loss occurs mostly at wound site, causing the surgically induced astigmatism [22] followed by the centre of the cornea.  The purpose of this study was to detect the effect of phacoemulsification surgery on corneal endothelium.

2. SUBJECTS AND METHODS

This prospective, observational study was done at King Fahd Hospital of the University (KFUH), Alkhobar, Saudi Arabia. After getting the Institutional Review Board (IRB) approval, the study was initiated at KFUH between April 2016 and August 2016. All patients with symptomatic age-related nuclear cataracts who don't have any one of the following exclusion criteria were included: history of previous ocular surgery, diagnosed cases of glaucoma, clinical evidence of retinal disease on dilated fundus examination, diagnosed cases of uveitis, patients with endothelial cell count which is less than 1500, intraoperative complications that were shown to impact the endothelial cells. After getting the patients' informed consents, a thorough assessment was done to each patient prior to surgery. The assessment included corrected and uncorrected distance visual acuity, slit-lamp assessment, Goldmann applanation tonometry, fundoscopy, intraocular lens calculation, and three times specular microscopy with average endothelial cell count of all of the readings was taken. Patients' have signed on informed consents for undergoing phacoemulsification surgery on the same day of evaluation. All eyes enrolled in the study have undergone phacoemulsification surgery procedure, in which a 2.2 mm temporal clear corneal incision under local anesthesia with a stop and shop technique to divide the nucleus were used. All the cases have been done using the endocapsular phaco technique, and the mechanical breakdown of the nucleus also done inside the bag whenever is needed. A WHITESTAR Signature phacoemulsification apparatus (Abbot Medical Optics) implementing the Elliptical cutting pattern was used for phacoemulsification in the capsular bag. Measurements of the intraoperative power (AVG %), elliptical motion (EFX), and ultrasound time (UST) were taken, with theatre staff documenting the duration of the surgical procedure. For each patient, one month after the phacoemulsification surgery , specular microscopy (CEM-530, NIDEK CO., LTD.) was performed three times. In both situations, the average endothelial cells density has been chosen. The phacodynamic parameters and surgery time were compared with the degree of endothelial cells loss, hexagonality and central corneal thickness. SPSS was used for the statistical analysis. A P value of less than 0.05 was considered statistically significant, and all data were expressed as the mean ± standard deviation.

3. RESULTS

 Of a study population of 175 patients, 96 were males (54.9%) and 79 were females (45.1%), with a mean age of 61.85 ± 9.12 years (range, 40 to 86 years).A total of 175 eyes (52.6% right eye and 47.4% left eye) were assessed in the study. Table 1 shows pre-operative and intra-operative parameters. Table 2 shows postoperative parameters. Corneal Endothelial cell density preoperative was 2569 ± 344 cells and postoperative was 2119 ± 508 postoperative. Central corneal thickness was 550 ± 37 micromtere preoperative and 564 ± 40 micrometre. Mean hexagonality preoperative was 67% ± 11% and postoperative mean hexagonality was 64% ± 15%.  Mean Corneal endothelial loss was 499 cells (17.4%) and mean difference in hexagonality was 2.5% and Central corneal thickness increased by a mean of 14.2 micrometer.  Age and the phacodynamic parameters used during phacoemulsification, namely, power (AVG %), ultrasound time, and elliptical motion showed a strong positive linear correlation with the degree of endothelial cell loss ( all P< .001) (Figure 1 a-c). The length of surgery also showed a positive relashionship with the amount of endothelial cell damage ( P < .001) (Figure 1 d).  None of the power, ultrasound time, surgery duration parameters have affected Hexagonality ( P 0.281, P 0.121, P 0.456 respectively). Only the Elleptical motion (P0.007). None of the parameters affected CCT significantly (P 0.209, P 0.117, P 0.919, P 0.175) for , power, ultrasound time, and elliptical motion and surgery duration respectively.

4. DISCUSSION

 It is clear that the phacoemulsification surgery affects the corneal endothelium despite various protective measures including the use of ocular viscoelastic devices [23] mechanical chopping forces and emulsification in the posterior chamber. The presence of good endothelial cell count before the surgery makes prominent endothelial cell damage causing corneal decompensation and pseudophakic bollous keratopathy unlikely. Such serious changes can occur even long time after phacoemulsification because of the physiological decrease in corneal endothelial cell count by 0.6% each year. Factors most likely to lead to more endothelial cell damage include phacoemulsification power use and length of surgery.

In this study, The percentage of endothelial cell loss was (17.4%) more than what was detected by Xing, et al. [22] (12.4%) and Walkow, et al. [24] (11.9%). Change is corneal endothelial cells morphology was significant by confocal microscopy [22] but not by specular microscopy as in our study. Change in corneal thickness was minimal (2.5%)  slightly more than 0.6% found by Salvi, et al. [25] and similar to vasavada et al (4.3%) and Sudeep et al (1.7%).  The factors which have significantly affected corneal endothelium were phaco power and ultrasound time, and elliptical motion. They were also significantly associated with corneal endothelal cell loss in O’Brien, et al. [26] and Walkow, et al. [24] studies although some studies failed to show this correlation [27-29].  Phaco power and amount of elliptical motion are related to the hardness of cataract and ultrasound time is affected by the hardness of cataract and the surgeon’s experience. Hardness of cataract and surgeon’s experience were also significantly associated with endothelial loss in Orski, et al. [19] study. The use of Transverse (elliptical) motion machine was thought to be safer but it was shown that it provides no superior protection for corneal endothelium than the tortional [30, 31] although this doesn’t apply to older tortional machines. Results also showed that as noted by previous studies, age was significantly associated with corneal endothelial loss as has been shown by Orski, et al. [19] although it is unmodifiable factor.

Fluidics and flow parameters were considered in other studies and showed that there was no significant effect on Central corneal thickness or corneal endothelial cell count [32, 33].

Funding: This study received no specific financial support.

Competing Interests: The authors declare that they have no competing interests.

Contributors/Acknowledgement: All authors contributed equally to the conception and design of the study.

REFERNCES

[1]          W. J. Stark, A. Sommer, and R. E. Smith, "Changing trends in intraocular lens implantation," Archives of Ophthalmology, vol. 107, p. 1441-1444, 1989. View at Google Scholar | View at Publisher

[2]          International Agency for Prevention of Blindness, World Blindness and its prevention. New York: Oxford University Press, 1980.

[3]          F. A. F. Mubarak, A. A. Al-Rajhi, O. M. Al-Omar, S. A. Al-Ghamdi, and J. Monzer, "Prevalence and causes of visual impairment and blindness in the South Western region of Saudi Arabia," International Ophthalmology, vol. 17, pp. 161-165, 1993. View at Google Scholar | View at Publisher

[4]          G. Ravalico, D. Tognetto, F. Baccara, and A. Lovisato, "Corneal endothelial protection by different viscoelasticsduring phacoemulsification," Journal of Cataract & Refractive Surgery, vol. 23, pp. 433–439, 1997. View at Google Scholar | View at Publisher

[5]          C. D. Kelman, "Phaco-emulsification and aspiration. A new technique of cataract removal: A preliminaryreport," American Journal of Ophthalmology, vol. 64, pp. 23–35, 1967. View at Google Scholar 

[6]          R. R. Bourne, D. C. Minassian, J. K. Dart, P. Rosen, S. Kaushal, and N. Wingate, "Effect of cataract surgery on the corneal endothelium: Modern phacoemulsification compared with extracapsular cataract surgery," Ophthalmology, vol. 111, pp. 679-685, 2004. View at Google Scholar | View at Publisher

[7]          H. B. Dick, T. Kohnen, F. K. Jacobi, and K. W. Jacobi, "Long-term endothelial cell loss following phacoemulsification through a temporal clear corneal incision," Journal of Cataract & Refractive Surgery, vol. 22, pp. 63–71, 1996. View at Google Scholar | View at Publisher

[8]          M. Kohlhaas, M. Klemm, J. Kammann, and G. Richard, "Endothelial cell loss secondary to two different phacoemulsification techniques," Ophthalmic Surgery, Lasers, vol. 29, pp. 890–895, 1998. View at Google Scholar 

[9]         N. Rosado-Adames and N. A. Afshari, "The changing fate of the corneal endothelium in cataract surgery," Current Opinion in Ophthalmology, vol. 23, pp. 3–6, 2012. View at Google Scholar | View at Publisher

[10]        D. M. Maurice, The cornea and sclera. In: Davson H, Editors. The Eye, 3rd ed. Orlando: Academic Press, 1984.

[11]        N. C. Joyce, D. L. Harris, and D. M. Mello, "Mechanisms of mitotic inhibition in corneal endothelium: Contact inhibition and TGF-beta2," Investigative Ophthalmology & Visual Science, vol. 43, pp. 2152-2159, 2002. View at Google Scholar 

[12]        N. C. Joyce, "Proliferative capacity of the corneal endothelium," Progress in Retinal and Eye Research, vol. 22, pp. 359–389, 2003. View at Google Scholar | View at Publisher

[13]        L. Rossetti, J. Chaudhuri, and K. Dickersin, "Medical prophylaxis and treatment of cystoid macular edema after cataract surgery. The results of a meta-analysis," Ophthalmology, vol. 105, pp. 397–405, 1998. View at Google Scholar | View at Publisher

[14]        R. A. Eiferman and E. L. Wilkins, "The effect of air on human corneal endothelium," American Journal of Ophthalmology, vol. 92, pp. 428-331, 1981. View at Google Scholar | View at Publisher

[15]        M. Packer, W. J. Fishkind, I. H. Fine, and R. S. Hoffman, "The physics of phaco: A review," Journal of Cataract & Refractive Surgery, vol. 31, pp. 424–431, 2005. View at Google Scholar 

[16]        K. Saito, K. Miyake, P. L. McNeil, and K. Kato, "Plasma membrane disruption underlies injury of the corneal endothelium by ultrasound," Experimental Eye Research, vol. 68, pp. 431–437, 1999. View at Google Scholar | View at Publisher

[17]        D. S. Hull, "Oxygen free radical and corneal endothelium," Transactions of the American Ophthalmological Society Journal, vol. 88, pp. 463–511, 1990. View at Google Scholar 

[18]        H. Takahashi, "Corneal endothelium and phacoemulsification," Cornea, vol. 35, pp. S3-S7, 2016. View at Google Scholar | View at Publisher

[19]        M. Orski, A. Synder, D. Pałenga-Pydyn, W. Omulecki, and M. Wilczyński, "The effect of the selected factors on corneal endothelial cell loss following phacoemulsification," Klinika Oczna, vol. 116, pp. 94-99, 2014. View at Google Scholar 

[20]        W. M. Bourne, L. R. Nelson, and D. O. Hodge, "Central corneal endothelial cell changes over a ten-year period," Investigative Ophthalmology & Visual Science, vol. 38, pp. 779–782, 1997. View at Google Scholar 

[21]        L. T. J. Chylack, M. C. Leske, D. McCarthy, and P. Khu, "Lens opacities classification system II (LOCS II)," Archives of Ophthalmology, vol. 107, pp. 991–997, 1989. View at Google Scholar | View at Publisher

[22]        D. Xing, Z. Guiqiu, W. Qing, Y. Xian, G. Ang, L. Jing, W. Qian, and X. Qiang, "Preliminary study of the association between corneal histocytological changes and surgically induced astigmatism after phacoemulsification," BMC Ophthalmology, vol. 14, p. 134, 2014. View at Google Scholar 

[23]        H. Tomomi and S. Kazuhisa, "Use of viscoelastic substance in ophthalmic surgery – focus on sodium hyaluronate," Clinical Ophthalmology, vol. 2, pp. 21–30, 2008. View at Google Scholar | View at Publisher

[24]        T. Walkow, N. Anders, and S. Klebe, "Endothelial cell loss after phacoemulsification: Relation to preoperative and intraoperative parameters," Journal of Cataract & Refractive Surgery, vol. 26, pp. 727–732, 2000. View at Google Scholar | View at Publisher

[25]        S. M. Salvi, T. K. Soong, B. V. Kumar, and N. R. Hawksworth, "Central corneal thickness changes after phacoemulsification cataract surgery," Journal of Cataract & Refractive Surgery, vol. 33, pp. 1426-1428, 2007. View at Google Scholar | View at Publisher

[26]        P. D. O’Brien, P. Fitzpatrick, D. J. Kilmartin, and S. Beatty, "Risk factors for endothelial cell loss after phacoemulsification surgery by a junior resident," Journal of Cataract & Refractive Surgery, vol. 30, pp. 839–843, 2004. View at Google Scholar | View at Publisher

[27]        A. Reuschel, H. Bogatsch, T. Barth, and R. Wiedemann, "Comparison of endothelial changes and power settings between torsional and longitudinal phacoemulsification," Journal of Cataract & Refractive Surgery, vol. 36, pp. 1855–1861, 2010. View at Google Scholar | View at Publisher

[28]        P. Kosrirukvongs, S. G. Slade, and R. G. Berkeley, "Corneal endothelial changes after divide and conquer versus chip and flip phacoemulsification," Journal of Cataract & Refractive Surgery, vol. 23, pp. 1006–1012, 1997. View at Google Scholar | View at Publisher

[29]        C. Zetterstrom and C. G. Laurell, " Comparison of endothelial cell loss and phacoemulsification energy during endocapsular phacoemulsification surgery," Journal of Cataract & Refractive Surgery, vol. 21, pp. 55–58, 1995. View at Google Scholar | View at Publisher

[30]        A. Ahmed and M. R. Maged, "Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision," Clinical Ophthalmology, vol. 7, pp. 55–61, 2013. View at Google Scholar | View at Publisher

[31]        F. S. Sorrentino, C. Bonifazzi, F. Parmeggiani, and P. Perri, "A pilot study to propose a harmscale, a new method to predict risk of harm to the corneal endothelium caused by longitudinal phacoemulsification, and the subsequent effect of endothelial damage on post operative visual acuity," PLoS One, vol. 11, p. e0146580, 2016. View at Google Scholar | View at Publisher

[32]        A. R. Vasavada, M. R. Praveen, V. A. Vasavada, V. A. Vasavada, S. M. Raj, and P. K. Asnani, "Impact of high and low aspiration parameters on postoperative outcomes of phacoemulsification: Randomized clinical trial," Journal of Cataract & Refractive Surgery, vol. 36, pp. 588–593, 2010. View at Google Scholar | View at Publisher

[33]        D. Sudeep, G. N. Soumya, K. K. Mathew, N. Somshekar, S. Rohit, and S. K. Bhujang, "Effect of fluidics on corneal endothelial cell density, central corneal thickness, and central macular thickness after phacoemulsification with torsional ultrasound," Indian Journal of Ophthalmology, vol. 63, pp. 641–644, 2015. View at Google Scholar | View at Publisher

Views and opinions expressed in this article are the views and opinions of the author(s), Journal of Cells shall not be responsible or answerable for any loss, damage or liability etc. caused in relation to/arising out of the use of the content.